Chapter 4 Command Reference
Introduction
Programming Guide 4-9
:CALCulate[1]:IMMediate:AUTO <Boolean>
Sets or queries whether post-processing (recalculation) will automatically occur
whenever any changes are made to the :CALCulate[1|2] subsystems.
With :CALC:IMM:AUTO set to OFF, :CALCulate[1|2] only produces new results
when new SENSe data is acquired or when the :CALCulate:IMMediate command
is received.
Once :CALC:IMM:AUTO is set to ON, the CALCulate[1|2] subsystems produce
new results when any CALCulate[1|2] command is processed, even when new
SENSe data is not being acquired.
This allows the user to make configuration changes in the CALCulate[1|2]
subsystems and immediately have new CALCulate[1|2] results on the same
SENSe data.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
This command affects all of the post-processing subsystems settings
(:CALC:IMM:AUTO).
Note that the Counter powers up with :CALC:IMM:AUTO set to ON, but
*RST sets it to OFF.
:CALCulate[1]:MATH Subtree
This subtree collects together the commands related to math (scale/offset)
processing. See the :TRACe subsystem for commands used to set the scale and
offset.
Many of these commands are query-only because the Counter has only one fixed
math operation.
:CALCulate[1]:MATH[:EXPRession]:CATalog?
Queries defined equation name.
The string SCALE_OFFSET is returned.
Query only.
Query Response
Comments
Query Response
Comments