Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide 2-27
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax Parameter Form Std/
New
Comments
:CALCulate3 (Cont.)
:LFILter
:LOWer
[:DATA]
:STATe
:UPPer
[:DATA]
:PATH?
<numeric_value> [HZ | S |
DEG]
<Boolean>
<numeric_value> [HZ | S |
DEG]
New
New
New
New
New
New
Std
Subtree. Limit FILter for statistics.
Subtree.
Sets the statistics filter lower limit.
Sets the statistics filter enable.
Subtree.
Sets the statistics filter upper limit.
Query only. Returns LFIL, AVER.
:CALibration
[:ALL]?
:COUNt?
:DATA
:SECurity
:CODE
:STATe
<arbitrary block>
<NRf>
<Boolean>, <NRf>
Std
Std
New
Std
New
New
New
Subsystem.
Query only. Causes an internal interpolator
self-calibration.
Query only. Returns value indicating number
of times the Counter has been calibrated.
Transfers the calibration data (input gain,
input offset, reference oscillator, and time
interval).
No query. Sets the calibration security code.
Enables or prevents calibration of the
Counter. Query returns security status. 0 =
unsecure; calibration allowed. 1 = secure;
calibration disallowed.
:CONFigure
Std See Measurement Instructions in this table.
:DIAGnostic
:CALibration
:INPut[1|2]
:GAIN
:AUTO
:OFFSet
:AUTO
:INTerpolator
:AUTO
ONCE | OFF
ONCE | OFF
ONCE | OFF | ON
Std
New
New
New
New
New
New
New
New
Subsystem.
Subtree.
Subtree. 1 | 2 specifies channel.
Subtree.
ONCE calibrates input gain.
Subtree.
ONCE calibrates input offset.
Subtree.
ONCE calibrates the interpolators.