HP 132A User Manual Page 220

  • Download
  • Add to my manuals
  • Print
  • Page
    / 330
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 219
Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-54 Programming Guide
This command disables math, statistics, and limit-testing.
If an absolute trigger level is not specified in the <parameters>, then when this
command executes, for functions other than Voltage Peaks (maximum,
minimum, peak-to-peak) or Totalize,
– auto-trigger is enabled,
– auto-trigger level(s) are set,
– auto-trigger is invoked on measurement channel(s).
:CONFigure?
Queries the function configured by the last :CONFigure or :MEASure query.
If the instrument state has changed through commands other than :CONFigure or
:MEASure query, the instrument will not track these changes, and the query
response will not reflect these changes.
A string of the form: <function> <parameters>[,<source_list>], omitting
the leading colon from the <function>.
The Rise Time, Fall Time, and Duty Cycle functions each have two ways of
being specified. This query response will use RTIM, FTIM, or DCYC to
indicate these functions.
The response is unaffected by *RST, recall, and [:SENS]:FUNC.
At power-on, this query generates an error and returns an empty string.
Refer to the sub-section in this section titled Descriptions of Measurement
Functions for descriptions of each measurement function.
Refer to Table 4-1 in this section for a summary of the <function>,
<parameters>, and <source_list> for each of the measurement functions.
Query Response
Comments
Page view 219
1 2 ... 215 216 217 218 219 220 221 222 223 224 225 ... 329 330

Comments to this Manuals

No comments