specifications
8753ET/ES
Network Analyzers
HP 8753ET and 8753ES Network Analyzers
30 kHz to 3 or 6 GHz
This document describes the performance
and features of the following products:
HP 8753ES S-parameter vector network analyzer
HP 8753ES Option 011 vector network analyzer without test set
HP 8753ET Transmission/reflection vector network analyzer
For more information about these analyzers, please see the following documents:
HP literature number
HP 8753ET and 8753ES Network Analyzers Brochure 5968-5159E
HP 8753ET and 8753ES Network Analyzers Configuration Guide 5968-5158E
Page view 0
1 2 3 4 5 6 ... 27 28

Summary of Contents

Page 1 - 30 kHz to 3 or 6 GHz

specifications8753ET/ESNetwork AnalyzersHP 8753ET and 8753ES Network Analyzers30 kHz to 3 or 6 GHzThis document describes the performanceand features

Page 2 - Table of contents

10HP 8753ES specificationsTest-port output characteristics7Frequency characteristicsRange30 kHz to 3 GHz Option 006 30 kHz to 6 GHzResolution 1 HzStab

Page 3

11Test port input characteristics (continued)External source mode1(CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements

Page 4 - System performance summaries

12HP 8753ES Option 011 specificationsTest port output characteristicsFrequency characteristicsRange300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolu

Page 5

13HP 8753ES Option 011 specifications Test port input characteristics (continued)Frequency offset mode3Frequency range 300 kHz to 3 GHzOption 006 300

Page 6

14HP 8753ET specificationsTest-port output characteristicsFrequency characteristicsRange300 kHz to 3 GHzOption 006 300 kHz to 6 GHzResolution 1 HzSta

Page 7

15HP 8753ET specificationsTest-port input characteristics (continued)External source mode1(CW time sweep only) Frequency range 300 kHz to 6 GHz R-cha

Page 8

16HP 8753ET/ES supplementalcharacteristicsMeasurementNumber of display channelsFour display channels available.Number of measurement channels Two prim

Page 9 - Type-N test ports

17HP 8753ET/ES supplementalcharacteristics (continued)Sweep triggerSet to either continuous, single, group sweep, or exter-nal trigger. Set external t

Page 10 - HP 8753ES specifications

18HP 8753ET/ES supplementalcharacteristics (continued)Trace functionsDisplay data Display current measurement data, memory data, orcurrent measuremen

Page 11 - 2. Typical performance

19HP 8753ET/ES supplementalcharacteristics (continued)Frequency offset on/offSets the RF source to be swept at a fixed offset fre-quency above the rec

Page 12

2Table of contentsDefinitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3System performance summaries . . . . . .

Page 13

20General characteristicsFront panel connectors HP 8753ES test ports (without Option 011)Connector type7 mm, precision Impedance 50 ohms (nominal) HP

Page 14 - HP 8753ET specifications

21Environmental characteristicsGeneral conditionsRFI and EMI susceptibility: defined by VDE 0730,CISPR Publication 11, and FCC Class B Standards.ESD (

Page 15

22Test set specifications for Option 011 HP 85046A/B S-parameter test setsThe HP 85046A/B S-parameter test sets provide thecapability to measure refle

Page 16 - HP 8753ET/ES supplemental

23HP 85047A S-parameter test setThe HP 85047A S-parameter test set provides thecapability to simultaneously measure the reflectionand transmission cha

Page 17

24AccessoriesCalibration kitsVector accuracy enhancement procedures require thatthe systematic errors of the measurement system becharacterized by mea

Page 18

25HP 85038 family of calibration kitsThe HP 85038A calibration kit contains open circuits,short circuits, and terminations with both male andfemale 7-

Page 19 - Remote programming

26Verification kitsMeasuring known devices other than the standardsused in calibration is an easy way to verify the properoperation of an HP 8753ES me

Page 20 - General characteristics

27HP 11852B 50-ohm to 75-ohm minimum loss pad(continued)Connectors50-ohm type-N (f) to75-ohm type-N (m)standard Option 004, 50-ohm type-N (m) to 75-oh

Page 21 - Environmental characteristics

28HP 8347A RF amplifierThe HP 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The HP 8347A provides leveled out

Page 22 - HP 85046A/B Specifications

3Definitions and test conditionsSpecifications describe the instrument's warrantedperformance after a half-hour warm-up and over thetemperature r

Page 23 - 4. Typical performance

4System performance summariesHP 8753ES (50-ohm systems) 7-mm test portsConfigurationNetwork analyzer HP 8753ES Standard and Option 006Calibration kit

Page 24 - Accessories

5System performance summariesHP 8753ES (50-ohm systems)Type-N test portsConfigurationNetwork analyzer HP 8753ESStandard and Option 006Calibration kit

Page 25

6System performance summariesHP 8753ES (50-ohm systems)3.5-mm test portsConfigurationNetwork analyzer HP 8753ESStandard and Option 006Calibration kit

Page 26 - Test-port return cables

7System performance summariesHP 8753ES (75-ohm systems)Type-N test portsConfigurationNetwork analyzer HP 8753ES Option 075Calibration kit HP 85036BTes

Page 27 - RF limiter

8System performance summariesHP 8753ES (75-ohm systems)Type-F test portsConfigurationNetwork analyzer HP 8753E Option 075Calibration kit HP 85039BTest

Page 28 - HP 8347A RF amplifier

9System performance summariesHP 8753ET (50-ohm systems)Type-N test portsConfigurationNetwork analyzer HP 8753ET Standard, Option 006 or Option 004Cali