709.ICC. Refer to your operating system and/or application documentation for information on where to store
these ICC proles on your computer and activate them.
Optional calibration tags
The following optional tags can be added to the core calibration tags to provide additional capabilities. These
optional tags must be placed before the <calibration> element; they cannot be nested inside it. None of these
tags are required to be in a StudioCal XML le.
Measurement averaging
This tag allows you to make and average multiple measurements. If you are using a colorimeter for
calibration measurements, you may want to make multiple measurements of lower grayscale levels to
compensate for the noise inherent in some colorimeters.
NOTE: As spectroradiometers have less low-grayscale noise, this option is not typically used when
measuring with those instruments.
The tag contains two attributes and is structured as follows:
<average_meas N="INTEGER" use_below_RGB="INTEGER"/>
N: The number of measurements to be taken and averaged. Integer values between 2 and 10 are supported.
use_below_RGB: The RGB grayscale value below which you wish to make and average multiple
measurements. A single 8-bit value between 1 and 255 is supported.
Set linear segment
This tag allows you to specify a grayscale value at which the gamma will switch from the calculated curve to a
pure linear transition to zero. This is designed to allow you to mimic the measured performance of some
projectors.
The tag contains one attribute and is structured as follows:
<set_linear_section value="INTEGER"/>
The value attribute supports 10-bit values between 1 and 1024.
Select Klein cal le
This tag is specic to the Klein K10 and K10-A and allows you to specify any cal le stored in the instrument
as the matrix used for measurements. The default cal le for the Z31x is 15 and that is the one calibrated at
Klein Instruments, but you may desire to generate your own matrix using ChromaSurf.
This tag contains one attribute and is structured as follows:
<klein_cal value="INTEGER"/>
The value attribute supports integer values from 0–89.
Contrast ratio limiting
In some workows—on-screen newsprint, for example—it is desirable to have a contrast ratio lower than the
display’s native contrast ratio. The display provides a tag for this workow. The tag has one argument and is
structured as follows:
<cr_set cr="150"/>
Optional calibration tags 77