HP 300i Service Manual

Browse online or download Service Manual for PC/workstation barebones HP 300i. HP 300i Service manual User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 164
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
October 22, 1998 3:00 pm DRAFT
+
PageWriter 100, 200 and 300 Series Cardiographs
M1770A, M1771A, and M1772A
Mobile Cart M1705B
Service Manual
+33DUW1R0
3ULQWHGLQ86$2FWREHU
(GLWLRQ
&RS\ULJKW
+HZOHWW3DFNDUG&RPSDQ\
3ULQWHGLQ86$
Page view 0
1 2 3 4 5 6 ... 163 164

Summary of Contents

Page 1 - Service Manual

October 22, 1998 3:00 pm DRAFT+PageWriter 100, 200 and 300 Series Cardiographs M1770A, M1771A, and M1772A Mobile Cart M1705B Service Manual+33D

Page 2 - 0Notice

xCleaning the Electrodes and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21Caring for the Battery . .

Page 3

The Printer Door Assembly5-105The Printer Door Assembly This section describes how to remove and replace the printer door assembly. Figure 5-5 Removin

Page 4 - Limitation of Warranty

The Printer Drive Assembly5-11Replacing the Printer Door To replace the printer door, perform the above procedure in reverse order. Be sure to orient

Page 5 - Printing History

The Printer Drive Assembly5-12Replacing the Printer Drive Assembly To replace the printer drive assembly: • If necessary, reinstall the gears. The gea

Page 6 - Safety Summary

The Printhead Assembly5-135The Printhead Assembly This section describes how to remove and replace the printhead assembly. Removing the Printhead Asse

Page 7

The Printhead Assembly5-148. Carefully slide the printhead assembly up and out of the printer well. Pull the two bundles of wires and the ground wire

Page 8 - 0Preface

The Printhead Assembly5-155. Push the lens through the mounting hole. 6. Finish removing the sensor by pulling the sensor through the mounting hole.

Page 9 - Contents

The Printhead Assembly5-16Figure 5-10 Replacing the Foam Block Figure 5-11 Positioning of the Printhead A. Printhead switch pin B. Printhead C. Leaf s

Page 10

The Control Board Assembly5-175The Control Board Assembly This section describes how to remove and replace the control board assembly. This procedure

Page 11

The Power Supply5-185The Power Supply This section explains how to remove and replace the power supply. Use an electrostatic wrist band or other appro

Page 12

The Power Supply5-195. Separate the power supply from the lower case. 6. Disconnect connector J2 from the power supply by grasping all the wires of

Page 13 - Introduction

xiIntegrated Front End Circuits/Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12Calibration . . .

Page 14

The Fuses5-205The Fuses To replace the AC fuses: 1. Unplug the cardiograph from AC power. 2. Turn the cardiograph bottom-side up. 3. Locate the tw

Page 15 - Specification Data

The Internal Cables5-215The Internal CablesCabling between the power supply board, the control board, the capacitor board, and the printhead has chang

Page 16

The Internal Cables5-2211. Hold the cables in this position while sliding the power supply board into its guide slots in the lower case. Be careful n

Page 17

The Storage and Transmission Board Assembly5-23Replacing the Printer CablesMore detailed instructions for replacing the printhead power cable and the

Page 18 - Parameter Specification

The Storage and Transmission Board Assembly5-244. When the storage and transmission board assembly has been pried out at least 0.25 inches, it should

Page 19

6-1Parts Lists 5Introduction This chapter contains the parts lists for the cardiograph and cart. The numbers on each figure or exploded view refer to

Page 20 - Options and Accessories

Ordering Information6-25Ordering Information You can order any of the parts listed through your local Hewlett-Packard Sales/Service Office listed belo

Page 21

Calling for Service6-35Calling for Service Here are the numbers you can use to call Hewlett-Packard Service for technical assistance.United States of

Page 22

Parts List6-45Parts ListAll references to AHA mean that the color code conforms to recommendations of the American Heart Association. All references t

Page 23

Parts List6-5Patient cables, power cords, batteries, and other accessories are not supplied with rebuilt units. ABZ Italian(Metr)M1770-68505M1770-6840

Page 24

xiiThe Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13Remo

Page 25 - Performance Verification

Parts List6-6Figure 6-1 Patient Cable Exploded Diagram

Page 26 - Extended Self-test

Parts List6-71Includes 2 and 3.Table 6-2 Patient Cable Parts List Ref. No. Description Part No. 1 Patient cable, AHA1M2461APatient cable, IEC1M2462A

Page 27

Parts List6-8Figure 6-2 Main Assembly Exploded Diagram

Page 28

Parts List6-9The printer door assembly does not include the door label. To get the label, order the Case label set for the language needed. Table 6-3

Page 29

Parts List6-10Figure 6-3 Bottom Case Exploded Diagram Figure 6-4 Storage and Transmission Board16

Page 30 - Prefix CND

Parts List6-111When replacing capacitor assembly M1770-60162 or cable assembly M1770-61604, both parts must be replaced with capacitor assembly M1770-

Page 31

Parts List6-12Figure 6-5 Bottom Case Exploded Diagram Figure 6-6 Storage and Transmission Board16

Page 32

Parts List6-131When replacing capacitor assembly M1770-60162 or cable assembly M1770-61604, both parts must be replaced with capacitor assembly M1770-

Page 33

Parts List6-14Figure 6-7 Bottom Case–External Exploded Diagram

Page 34

Parts List6-15The battery door assembly does not include the door label. To get the label, order the case label set for the language needed. Table 6-8

Page 35

1-1Introduction This chapter introduces you to the M1770A PageWriter 200i/300pi, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs and lis

Page 36 - Printer Test

Parts List6-16Figure 6-8 Upper Case Exploded Diagram

Page 37 - Keyboard Tests

Parts List6-171M1770-61670 replaces M1770-61606Table 6-9 Upper Case Assembly Parts List Ref. No. Description Part No. 1 Gear plate and motor assembl

Page 38 - ECG Simulation

Parts List6-18Figure 6-9 M1770A/M1771A Keyboard Exploded Diagram

Page 39

Parts List6-191Complete assembly of items 1-82Items 2a, 2b, and 3 are not individually replaceable. These parts require factory assembly. Refer to ite

Page 40

Parts List6-20Figure 6-10 M1772A Keyboard Exploded Diagram

Page 41 - Resetting the PageWriter 100

Parts List6-211Complete assembly of items 1-32Items 2a, 2b, and 3 are not individually replaceable. These part require factory assembly. Refer to item

Page 42 - Preventive Maintenance

Parts List6-22Figure 6-11 M1705B Cart Exploded Diagram.

Page 43

Parts List6-23Table 6-12 M1705B Cart Parts List Ref. No. Description Part No. Cart assembly M1705-695001 Retainer cap (2 req’d) 0590-19772 Screw, tp

Page 44

Parts List6-24

Page 45

Parts List6-25

Page 46 - Caring for the Battery

PageWriter 100, 200, 200i, and 300pi Series Cardiographs1-2Figure 1-1 M1770A/M1771A PageWriter 300pi/200i/200 Cardiograph.The PageWriter 100, 200, a

Page 49

Parts List6-28

Page 50

Parts List6-29

Page 51

Parts List6-30

Page 52

Parts List6-31

Page 54 - Circuit Descriptions

A-1Connector Pin AssignmentsThe tables in this appendix list the pin assignments for each connector on the CPU assembly and the power supply boards. F

Page 55 - The Patient Cable

CPU Assembly ConnectorsA-2The connector listed in this table is shared by the keyboard and the preview screen. Pins 1–11 serve the keyboard, while pin

Page 56 - CPU Assembly

CPU Assembly ConnectorsA-3Signal Definition+5KBD 5 volt supply for keyboard assembly. ACONLED AC on LED. Drives the AC indicator LED on the keyboard.

Page 57 - CPU/System Oscillator

Inquiries1-3M1705B Cart The optional M1705B Cart provides mobility for the PageWriter 100, 200, and 300 family of cardiographs. The large wheels make

Page 58 - 2Q6WDQGE\

CPU Assembly ConnectorsA-4 Table A-2 M1770A/71A Keyboard Connector J102 to LCD Module Pin Signal Pin Signal 1 DF 9 DD22 Reserved 10 DD33ground11+5KB

Page 59 - Printer Gate Array/SRAM

CPU Assembly ConnectorsA-5Table A-3 Control Board J201 to M1772A Keyboard Connector J101 Pin Signal Pin Signal Pin Signal 1 ground 15 ROWSA_1 28 LED11

Page 60 - System Expansion Connector

CPU Assembly ConnectorsA-6Signal Definition+5KBD 5 volt supply for keyboard assembly. ACONLED AC on LED. Drives the AC indicator LED on the keyboard.

Page 61 - Power Supply

CPU Assembly ConnectorsA-7 Table A-4 Control Board Connector J304 to Printhead Logic Pin Signal Pin Signal 1 DATA_OUT 8 nSTROBE32 FSW_+5V 9 nSTROBE4

Page 62

CPU Assembly ConnectorsA-8Table A-5 Control Board Connector J303 to Paper Sensor Pin Signal Pin Signal 1 DOOR 4 GROUND2 GROUND 5 DETECTOR3 EMITTERSi

Page 63

CPU Assembly ConnectorsA-9 Table A-7 Power Supply Connector J3 to Control Board Connector J301 Pin Signal Pin Signal 1 ENBL5V 8 PPWRON2 GROUND 9 GR

Page 64 - Defibrillator Protection

CPU Assembly ConnectorsA-10Table A-8 Power Supply Connector J2 to Control Board J302 Pin Signal Pin Signal 1 GROUND 9 +5V2 GROUND 10 +5V3 GROUND 11

Page 65

CPU Assembly ConnectorsA-11Although AHA designations are used as signal names, IEC designations are more accurate descriptors. Table A-10 Control Boar

Page 66 - ROM-X1, ROM-X2

CPU Assembly ConnectorsA-12Table A-11 System Expansion Connector J12 Pin Row A Row B Row C Pin Row A Row B Row C 1 GROUND GROUND GROUND 17 A2

Page 67

CPU Assembly ConnectorsA-13FEDATA Serial data from front end. FEPWR Front end transformer drive control. FSW_+5V Filtered switched 5 V. GPIO0-GPIO3 Ge

Page 68

Specification Data1-41Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.2Meets or exceeds AAMI EC11-1991 standard f

Page 69 - Troubleshooting

CPU Assembly ConnectorsA-14 Table A-12 Power Supply J1 to AC Input Cable Pin Signal 1L12 EARTH_GND3L2

Page 70 - Test Tools

IndexIndex - 1AA1, A-14A16, A-12ACfuse holders, 5-20ACON, 4-19ACON signal, A-10, A-12ACONLED signal, A-3, A-6AD1, A-12ALE signal, A-12Bbattery, 2-22pr

Page 71 - Messages

IndexIndex - 2optical paper sensor removal, 5-14optionspower cords, 1-9options, cardiograph, 1-8, 1-12country, 1-8ordering information, parts, 6-1PPag

Page 72 - Troubleshooting Tables

IndexIndex - 3LED5, A-6LED6, A-6signalsA1, A-14A16, A-12ACON, A-10, A-12AD1, A-12ALE, A-12CLK2A, A-12CP, A-14D0, A-12DATA, A-8DATA_OUT, A-7DD0, A-4DET

Page 73

IndexIndex - 4

Page 74

Specification Data1-51Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Meets frequency response

Page 75

Specification Data1-61Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.2Meets or exceeds IEC 601-2-25 Internationa

Page 76

Specification Data1-71Described in the European Committee for Standardization, Standard Communications Protocol - Computer - Assisted Electrocardiogra

Page 77

Noticeii0Notice The information in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to thi

Page 78 - Chart Speed

Options and Accessories1-81Options and Accessories PageWriter 100, 200, 200i, and 300pi Cardiographs These tables list the options and accessories ava

Page 79

Options and Accessories1-91In Table 1-9 an “A” in the Printer column refers to 8.5 x 11-in. paper; “A4” refers to 210 x 297-mm paper2See Table 1-11 fo

Page 80

Options and Accessories1-10Table 1-10 PageWriter 300pi, 200/200i and 100 Series Documentation Part Numbers Document Title Language Part NumberPageWri

Page 81

Options and Accessories1-11PageWriter 300pi Predictive Instruments Quick Reference Guide (M1770A with Serial # Prefix CND)English M1770-92800French M1

Page 82

Options and Accessories1-12Standard Accessories Accessories included are based on model number and localization. • 200 sheets of z-fold paper-English

Page 83

2-1Performance Verification and Maintenance1Introduction This chapter describes how to verify the cardiograph’s performance, explains preventive maint

Page 84

Performance Verification2-2Visual Inspection and Power On Self TestBefore beginning the inspection, press the key to put the cardiograph in Standby

Page 85

Performance Verification2-3When the tests are completed, the results are also printed out in a pass/fail format on the left-hand side of the first pag

Page 86 - Problem Corrective Action

Performance Verification2-4Internal Circuitry Testing of PageWriter 200/200i/300pi These tests are performed on the internal circuitry: •Memory • Gate

Page 87 - Testing the Power Supply

Performance Verification2-5If your cardiograph is equipped with Options #A05 or #A10, do not turn it off during Test C or D. If you turn off the cardi

Page 88

NoticeiiiLike all electronic devices, this cardiograph is susceptible to electrostatic discharge (ESD). Electrostatic discharge typically occurs when

Page 89

Performance Verification2-6Table 2-5 ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial NumberPrefix CNA, CNB, or CNC,3807A

Page 90

Performance Verification2-7Table 2-7 ROM Replacement Part Numbers for PageWriter Cardiographs M1772A Serial NumberPrefix CNA, CNB, or CNC,3807A or ear

Page 91 - Removal and Replacement

Performance Verification2-8Non-Volatile Random Access Memory (test 6) . Each 2-byte location of NVRAM is written and read with a pseudo-random pattern

Page 92 - The Battery

Performance Verification2-9Flash Memory Tests (tests C and D) . The flash memory tests verify that the flash memory on the optional expansion board ca

Page 93 - A Battery door

Performance Verification2-10Figure 2-1 An Extended Self-Test Report Example A. Test Results B. Character Set C. Timing Tics D. Diagonal Lines E. Stepp

Page 94 - The Keyboard Assembly

Performance Verification2-11How to Read the Extended Self-Test Report Each printed page of the extended self-test contains the following information:

Page 95

Performance Verification2-12Display Test (PageWriter 200/200i/300pi only) The display test shows test patterns on the display. Each of the following p

Page 96

Performance Verification2-13Keyboard Tests The keyboard tests are not automatically performed during the extended self-test. These must be manually pe

Page 97 - Replacing the Keyboard

Performance Verification2-14ECG Simulation Taking an ECG using a 12-lead ECG simulator allows you to verify areas of operation that the extended self-

Page 98 - The Top Cover Assembly

Performance Verification2-15Figure 2-2 12-Lead ECG from ECG Simulator (200/200i/300pi only).

Page 99

NoticeivWarranty Hewlett-Packard warrants this medical product against defects in materials and workmanship for a period of three years in certain geo

Page 100 - The Printer Door Assembly

Changing the Default Operating Language (200/200i/300pi Only)2-162Changing the Default Operating Language (200/200i/300pi Only) The default operating

Page 101 - The Printer Drive Assembly

Resetting the Cardiograph to the Factory Default State2-172Resetting the Cardiograph to the Factory Default State Resetting the cardiograph sets the A

Page 102

Preventive Maintenance2-182Preventive Maintenance Routine maintenance of the cardiograph consists of cleaning and inspection. This should be done as n

Page 103 - The Printhead Assembly

Preventive Maintenance2-19Cleaning the Keyboard Overlay 1. Carefully insert a small flat-bladed screwdriver into the notches at the front edge of the

Page 104

Preventive Maintenance2-20Cleaning the Digital Array Printhead and Paper Sensor If you use Hewlett-Packard recording paper, you should not have to cle

Page 105

Preventive Maintenance2-212. Wipe the printhead with a foam swab dipped in 90% isopropyl alcohol. Scrub until all visible residue is removed. 3. Dry t

Page 106

Caring for the Battery2-222Caring for the Battery The battery must be installed for proper operation of the cardiograph. Even if the cardiograph is pl

Page 107 - The Control Board Assembly

Caring for the Battery2-23Safety Tests The safety tests listed below are performed at the time of manufacture to assure compliance with these standard

Page 108 - The Power Supply

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix2-242PageWriter 100, 200/200i, and 300pi Series Performance Verification Mat

Page 109

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix2-25Extended Self Test Function Tests (F) Run Extended Self Test; verify:•RO

Page 110 - The Fuses

NoticevPrinting History October 1994 Edition 1May 1996 Edition 2October 1998 Edition 3

Page 111 - The Internal Cables

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix2-26Printer Test (P)for M1772Verify:• straight diagonal lines• even spacing

Page 112

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix2-27System Safety (S)Use Safety AnalyzerCheck:Chassis-to-Ground ResistanceGr

Page 113 - Replacing the Printer Cables

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix2-28

Page 114

3-1Theory of OperationThis chapter contains an overview of the cardiograph operation, and circuit descriptions of the major subassemblies. Except as n

Page 115 - Parts Lists

Circuit Descriptions3-2Power-off The power-off sequence is software controlled. The power-off sequence is initiated when the key is pressed, battery

Page 116 - Ordering Information

The Patient Cable3-3Figure 3-1 Simplified System Block Diagram 3The Patient Cable The patient cable used in the PageWriter 100/200/200i/300pi family i

Page 117 - Calling for Service

CPU Assembly3-43CPU Assembly This discussion of the CPU assembly covers the following circuits: •CPU • System Gate Array • Serial EEROM • System Memor

Page 118 - Parts List

CPU Assembly3-5CPU/System Oscillator The CPU (Central Processing Unit) is the processing engine for the instrument. The microprocessor used is the Int

Page 119

CPU Assembly3-6DMA for ECG Data The System Gate Array receives data from the ECG Front End in serial form. When 16 bits have been received, the System

Page 120

CPU Assembly3-7NVRAM Interface The NVRAM contains 1K bits of non-volatile storage for the cardiograph configuration. Keyboard Interface The passive Ke

Page 121

Safety Summaryvi0Safety Summary Safety Symbols Marked on the Cardiograph The following symbols are used on the cardiograph or the cart: Please see Cha

Page 122

CPU Assembly3-8timing and the printer RAM, regulates motor speed, monitors printhead temperature, monitors the battery condition, and provides the int

Page 123

Power Supply3-93Power Supply The Power Supply connects the instrument to the AC line to charge the battery and power the circuitry. It conditions and

Page 124

Power Supply3-10Rectifier/Filter This circuitry rectifies and filters the AC voltage to produce a DC voltage for the rest of the power supply. Switch-

Page 125

ECG Front End3-113ECG Front End The ECG Front End provides the interface between the system and the patient. It electrically isolates the patient to p

Page 126

ECG Front End3-12Opto-Isolators Opto-Isolators pass the digital sample data from the patient-connected circuitry to the rest of the instrument, provid

Page 127 - Language Part Number

Storage and Transmission (Option #A05 0nly)3-133Storage and Transmission (Option #A05 0nly)The Option #A05 PC assembly plugs into the cardiograph expa

Page 128

Storage and Transmission (Option #A05 0nly)3-14Figure 3-6 Option #A05 PC Assembly Block Diagram M177XA Expansion ConnectorThe expansion connector on

Page 129

Storage and Transmission (Option #A05 0nly)3-15Flash1, Flash2FLASH1 and FLASH2 are 1M x 16 bit flash ROMs. The gate array provides control and timing

Page 130

Storage and Transmission (Option #A05 0nly)3-16

Page 131

4-1Troubleshooting3Introduction This chapter provides information for localizing cardiograph problems to the subassembly level. This information is de

Page 132

Conventions Used in This Manualvii0Conventions Used in This Manual Warning statements describe conditions or actions that can result in personal injur

Page 133

Test Tools4-23Test ToolsTwo troubleshooting tools are provided with each instrument:• patient cable test tool (M1770-87908)• front end and transmissio

Page 134

The Error and Event Logging4-33The Error and Event Logging The last error that was displayed can be printed by starting the Extended Self-test. The er

Page 135

Troubleshooting Tables4-4PageWriter 100 Error Message Communication The PageWriter 100 communicates the following error messages with these correspond

Page 136

Troubleshooting Tables4-5Before removing or inserting any board or connector, make sure AC power is off and the battery is removed. The error codes li

Page 137 - × 2 (2 req’d)

Troubleshooting Tables4-62Table 4-2 Power Supply and Battery Symptom Possible Cause Corrective Action AC indicator not lit. Power cord unplugged.AC

Page 138

Troubleshooting Tables4-7Low Battery light or Low Battery message remains on during extended self-test.Low battery or defective battery.Battery fuse i

Page 139

Troubleshooting Tables4-8Remember that most noise results from poor patient-electrode connections or poor quality electrodes. If noise persists after

Page 140

Troubleshooting Tables4-9You can use the patient cable test tool and the ECG front end test tool to determine whether a problem is caused by a poor pa

Page 141

Troubleshooting Tables4-10Message: Printer door open or the current LED is flash-ing.Door is ajar.A printer cable is not fully seated.Printhead cont

Page 142

Troubleshooting Tables4-11Printed data are garbled. A printer cable is not fully seated.Defective printhead.Defective component in data path.Reseat co

Page 143

Prefaceviii0PrefaceThis manual contains service information for the Hewlett-Packard M1770A pageWriter 300pi, PageWriter 200i, M1771A PageWriter 200, a

Page 144

Troubleshooting Tables4-12Table 4-7 Storage (Option #A05 0nly)Message Possible Cause Possible SolutionsECG too noisy to store(This message appears bri

Page 145

Troubleshooting Tables4-13Table 4-8 Transmission (Option #A05 only)Message Possible Cause Possible SolutionsTelephone busy, waiting to redial Busy tel

Page 146

Troubleshooting Tables4-14No modem at remote site The local modem could not complete the connection. The most likely causes are:1. Telephone cable dis

Page 147 - Connector Pin Assignments

Troubleshooting Tables4-15In addition to the error codes listed above, the following event codes are displayed on the extended self-test report for ca

Page 148 - CPU Assembly Connectors

Troubleshooting Tables4-165160, sub codes 10 and 11Check cable. (This message appears when the trans-mission type is Direct or Direct SCP)Check modem

Page 149 - On-Standby

Troubleshooting Tables4-179030-9035, 9042-9045, 9050, 9051, 9069-9085 Direct transmission protocol error. Retry transmission.9038 A non-specific error

Page 150 - Pin Signal Pin Signal

Troubleshooting Tables4-18Table 4-10 Recorder ProblemsProblem Corrective ActionIs the cardiograph turned on? The LCD screen should be on. Use the shif

Page 151

Testing the Power Supply4-194Testing the Power Supply Use this procedure to check the voltages produced by the power supply. The cardiograph must be c

Page 152 - On/Standby

Testing the Power Supply4-208. Turn the cardiograph On by pressing . 9. Measure SW+5V between ground and the point marked SW+5V. It should read +5

Page 153

Testing the Power Supply4-21Figure 4-2 Voltage Test Locations

Page 154

ixContentsNotice. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 155

Testing the Power Supply4-22

Page 156

5-1Removal and Replacement4Introduction This chapter contains procedures for removing and replacing these cardiograph subassemblies: • battery • key

Page 157

The Battery5-2The directions in this chapter generally assume you are facing the front of the cardiograph as you work. Thus, for example, the “left si

Page 158

The Battery5-3Figure 5-1 Removing the Battery Door.A Battery door

Page 159

The Keyboard Assembly5-45The Keyboard Assembly This section explains how to remove and replace the keyboard assembly. The keyboard procedures apply to

Page 160 - Pin Signal

The Keyboard Assembly5-5Figure 5-2 Removing the Keyboard Overlay A. Overlay notches 5. Using two screwdrivers, place the blade of one screwdriver on

Page 161 - Index - 1

The Keyboard Assembly5-66. Carefully lift the assembly and tilt it until the keyboard and control board are accessi-ble. 7. Disconnect the keyboard

Page 162 - Index - 2

The Keyboard Assembly5-7Replacing the Keyboard To replace the keyboard, perform the following steps: 1. Connect the green wire to the spade connector

Page 163 - Index - 3

The Top Cover Assembly5-85The Top Cover Assembly This section describes how to remove and replace the top cover assembly. The top cover assembly inclu

Page 164 - Index - 4

The Top Cover Assembly5-9The capacitors on the connection assembly next to the control board store hazardous amounts of energy during AC and battery o

Comments to this Manuals

No comments