Hp E1368A User Manual Page 59

  • Download
  • Add to my manuals
  • Print
  • Page
    / 88
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 58
[ROUTe:]
The ROUTe subsystem controls switching and scanning operations for the
Microwave Switch modules in a switchbox.
Subsystem Syntax [ROUTe:]
CLOSe <
channel_list>
CLOSe? <
channel_list>
OPEN <
channel_list>
OPEN? <
channel_list>
SCAN <
channel_list>
:MODE NONE | VOLT
:MODE?
CLOSe [ROUTe:]CLOSe <channel_list> closes the Microwave Switch channels
specified in the channel_list. The channel_list is in the form (@ccnn)
where cc = card number (00-99) and nn = channel number (00-04).
Parameters
Parameter Name Parameter Type Range of Values Default Value
channel_list
numeric cc00 - cc04 N/A
Comments Closing Channels: To close:
a single channel, use [ROUTe:]CLOSe (@ccnn);
multiple channels, use [ROUTe:]CLOSe (@ccnn,ccnn);
sequential channels, use [ROUTe:]CLOSe (@ccnn:ccnn);
a group of sequential channels, use
[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn);
or any combination.
Closure order for multiple channels with a single command is not
guaranteed.
Related Commands: [ROUTe:]OPEN, [ROUTe:]CLOSe?
*RST Condition: All channels are open.
Example Closing Microwave Switch Channels
This example closes channels 100 and 202 of a two-module switchbox (card
numbers 01-02).
CLOS (@100,202) 100 closes channel 00 of card #1;
202 closes channel 02 of card #2.
58 HP E1368A/69A/70A Command Reference Chapter 4