Scanning Using Bus Triggers
This example uses bus triggering (
TRIG:SOUR BUS) to scan channels 00
through 02 of a single-module switchbox. The trigger source to advance the
scan is the
*TRG command (as set with TRIGger:SOURce BUS). When
INIT is executed, the scan is started and channel 00 is closed. Then, each
*TRG command advances the scan to the next channel.
TRIG:SOUR BUS Select interface (BUS) triggering.
SCAN (@100:102) Scan channels 00 through 02.
INIT Start scan, close channel 00.
*TRG Advances scan using BUS
triggering.
increment loop Increment loop count.
:SOURce? TRIGger:SOURce? returns the current trigger source for the switchbox.
Command returns either
BUS, EXT, HOLD, or IMM for trigger sources BUS,
EXTernal, HOLD, or IMMediate, respectively.
Example Query the Trigger Source
TRIG:SOUR EXT Sets trigger source to external.
TRIG:SOUR? Query trigger source.
Chapter 4 HP E1368A/69A/70A Command Reference 69