HP 34970A User Manual Page 111

  • Download
  • Add to my manuals
  • Print
  • Page
    / 429
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 110
Thermistor Measurements
To connect a thermistor to the module’s screw terminals, see page 21.
The instrument supports 2.2 k
(44004), 5 k(44007), and 10 k
(44006) thermistors.
Front-Panel Operation: To select the thermistor function for the
active channel, choose the following items.
TEMPERATURE , THERMISTOR
To select the thermistor type for the active channel, choose from the
following items.
TYPE 2.2 KOHM , TYPE 5 KOHM , TYPE 10 KOHM
Remote Interface Operation: You can use the MEASure? or
CONFigure command to select the probe type and thermistor type.
For example, the following statement configures channel 301 for
measurements of a 5 k
thermistor:
CONF:TEMP THER,5000,(
@301)
You can also use the SENSe command to select the probe type and
thermistor type. For example, the following statement configures
channel 103 for measurements of a 10 k
thermistor:
SENS:TEMP:TRAN:THERM:TYPE 10000,(
@103)
Chapter 4 Features and Functions
Temperature Measurement Configuration
112
Page view 110