• Some checkpoints can use the
same result data. Method
Validation Pack provides a wiz-
ard to define multiple usage of
analytical results as shown in
figure 23.
Method Validation Pack also trans-
fers these sequences to the
ChemStation for data acquisition
and data analysis. The sequences
can load into an Agilent Method
Validation Pack ChemStation sys-
tem and run similar to other
sequences in the Agilent
ChemStation. The results of the
method validation sequences are
stored automatically in the
ChemStore database. Each valida-
tion corresponds to a separate
ChemStore study.
34
Test execution
Having completed the validation
configuration, the experiments
can be executed on an Agilent
ChemStation Plus system. Method
validation pack transfers the theo-
retical validation planning into a
list of analytical runs by creating
one or multiple ChemStation
method validation sequences
(MVS). The user has two options
during sequence generation:
• The validation can be trans-
ferred into one sequence for the
complete validation or in multi-
ple validations experiments
where each component creates
an individual sequence as
shown in figure 22.
Figure 22
Options for generation of Method
Validation sequence: One sequence for
each component or one sequence for
the entire validation
Figure 23
Wizard for multiple usage of checkpoint results