*This bulletin supersedes TB 9-6625-2049-35, 15 January 1996.
*TB 9-6625-2049-35
SUPERSEDED COPY DATED 15 JANUARY 1996
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
SYNTHESIZER/FUNCTION GENERATOR
HEWLETT-PACKARD
MODELS 3325A( ) AND 3325B( )
Headquarters, Department of the Army, Washington, DC
2 January 2001
Approved for public release; distribution is unlimited
Paragraph
Page
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification............................
1 2
Forms, records, and reports................................
2 2
Calibration description .......................................
3 2
II. EQUIPMENT REQUIREMENTS
Equipment required............................................
4 4
Accessories required ...........................................
5 5
III CALIBRATION PROCESS
Preliminary instructions.....................................
6 6
Equipment setup.................................................
7 6
Harmonic distortion............................................
8 7
Amplitude modulation distortion........................
9 9
Square wave characteristics...............................
10 9
Ramp retrace.......................................................
11 10
Frequency accuracy............................................
12 11
Phase increment .................................................
13 13
Sine wave amplitude and flatness......................
14 14
Square wave amplitude and flatness..................
15 16
REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS
You can help improve this publication. If you find any mistakes or if you
know of a way to improve the procedure, please let us know. Mail your letter
or DA Form 2028 to: Commander, U. S. Army Aviation and Missile
Command, ATTN: AMSAM-MMC-LS-LP, Redstone Arsenal, AL 35898-5230.
A reply will be furnished to you. You may also send in your comments
electronically to our e-mail address: ls-[email protected] or by FAX (256)
842-6546/DSN 788-6546
Page view 0
1 2 3 4 5 6 ... 22 23

Summary of Contents

Page 1 - *TB 9-6625-2049-35

__________ *This bulletin supersedes TB 9-6625-2049-35, 15 January 1996. *TB 9-6625-2049-35 SUPERSEDED COPY DATED 15 JANUARY 1996 DEPARTMENT OF THE

Page 2

TB 9-6625-2049-35 10 (a) SIGNAL off (option 002 not lit). (b) FUNCTION square wave. (c) ENTRY FREQ. (d) 1 MHz. (e) ENTRY AMPTD. (f) 1 V RMS. (2

Page 3 - TB 9-6625-2049-35

TB 9-6625-2049-35 11 (4) Set oscilloscope controls to measure ramp retrace time from the 90 to 10 percent points. Ramp retrace time will be ≤3 µs.

Page 4 - Not verified below 50 Hz

TB 9-6625-2049-35 12 (15) Set frequency counter controls to measure frequency. If frequency counter does not indicate between 59.99970 and 60.00030

Page 5

TB 9-6625-2049-35 13 (12) Adjust A9R7 Fine Adj. (fig. 3) and A9 Coarse Adj. (fig. 3) for a minimum indication on frequency difference meter. (13) Di

Page 6

TB 9-6625-2049-35 14 (6) Adjust frequency counter sample rate control to HOLD and press RESET pushbutton. Record frequency counter indication. (7)

Page 7

TB 9-6625-2049-35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495

Page 8

TB 9-6625-2049-35 16 Table 4. Flatness Test instrument and measuring receiver settings (MHz) 1 2 4 6 8 10 12 14 16 18 20 b. Adjustments. No adjus

Page 9

TB 9-6625-2049-35 17 (a) 1 kHz. (b) ENTRY AMPTD. (c) 1 V RMS. (d) AMPTD CAL. (7) Multimeter will indicate between 0.99 and 1.01 V ac. (8) Disco

Page 10

TB 9-6625-2049-35 18 (6) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and 3.032 V ac. (7) Press keys and enter values using

Page 11

TB 9-6625-2049-35 19 Table 5. Dc Offset - Continued Test instrument DC OFFSET Multimeter indications (V dc) settings Min Max 14.99 mV 0.01493 0

Page 12

TB 9-6625-2049-35 2 Paragraph Page SECTION III. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ...

Page 13

TB 9-6625-2049-35 20 (4) Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.35 and -4.65 V dc. (5) Press ENTRY FREQ key and enter

Page 14

TB 9-6625-2049-35 21 (a) ENTRY FREQ. (b) 100 kHz. (c) ENTRY AMPTD. (d) 10 V RMS. (e) AMPTD CAL. (5) Multimeter will indicate between 9 and 11 V

Page 15

TB 9-6625-2049-35 22 Figure 5. Power supply assembly A2 – test instrument top view. (5) Press presently active FUNCTION key to remove ac output and

Page 16

TB 9-6625-2049-35 By Order of the Secretary of the Army: Official: ERIC K. SHINSEKI General, United States Army

Page 17

TB 9-6625-2049-35 3 Table 1. Calibration Description Test instrument parameters Performance specifications Harmonic distortion1 (relative to fundam

Page 18

TB 9-6625-2049-35 4 Table 1. Calibration Description - Continued Test instrument parameters Performance specifications Amplitude accuracy with dc of

Page 19

TB 9-6625-2049-35 5 5. Accessories Required. The accessories required for this calibration are common usage accessories, issued as indicated in parag

Page 20

TB 9-6625-2049-35 6 SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a. The instructions outlined in paragraphs 6 and 7 are preparatory to

Page 21

TB 9-6625-2049-35 7 8. Harmonic Distortion a. Performance Check (1) Connect TI EXT REF IN 1, 10 MHz to spectrum analyzer 10 MHz REF IN/OUT. (2) Con

Page 22

TB 9-6625-2049-35 8 (a) 50 Hz. (b) ENTRY AMPTD. (c) 10 VOLT. (13) Set audio analyzer to measure distortion in dB. Audio analyzer indication will

Page 23 - PIN: 052316-000

TB 9-6625-2049-35 9 9. Amplitude Modulation Distortion a. Performance Check (1) Press keys and enter values using DATA keys as listed in (a) through