HP 16500A User Manual Page 230

  • Download
  • Add to my manuals
  • Print
  • Page
    / 272
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 229
8086 or 8088 CPU Package: 40-pin DIP
Accessories Required: HP 10305B Preprocessor
HP 10269C General Purpose Probe Interface
Maximum Clock Speed: 10 MHz clock input (at CLK)
Signal Line Loading: Maximum of two 74ALS TTL loads + 40 pF on any
line
Microprocessor Cycles Identified: Memory read/write
I/O read/write
Code fetch
Interrupt acknowledge
Halt acknowledge
Transfer to 8087 or 8089
co-processors
Additional Capabilities: The 8086 or 8088 can be operating in
Minimum or Maximum modes. The logic
analyzer can capture all bus cycles (including
prefetches) or can capture only executed
instructions. To capture only executed
instructions, the 8086 the 8086 or 8088 must
be operating in the Maximum Mode.
Maximum Power Required: 1.0 A at + 5 Vdc, supplied by the logic
analyzer
Number of Probes Used: Three 16-channel probes
Microprocessor Specific Measurements HP 16510B
14 - 6 Front-panel Reference
Page view 229
1 2 ... 225 226 227 228 229 230 231 232 233 234 235 ... 271 272

Comments to this Manuals

No comments