What Am I
Going to
Measure?
After configuring the logic analyzer and hooking it up to your circuit
under test, you will be measuring the time (x) from when the RAS goes
low to when the CAS goes high, as shown below.
How Do I
Configure the
Logic Analyzer?
In order to make this timing measurement, you must configure the logic
analyzer as a timing analyzer. By following these steps you will configure
Analyzer 1 as the timing analyzer.
If you are in the State/Timing E Configuration menu you are in the right
place and you can start with step 2; otherwise, start with step 1.
1. Touch the field in the upper left corner of the display and select
State/Timing E.
2. In the State/Timing E Configuration menu, change Analyzer 1
type to Timing. If Analyzer 1 is already a timing analyzer, go on to
step 3.
a. Touch the field Type: _________ .
b. When the pop-up appears, touch Timing.
Figure 7-1. RAS and CAS Signals
Using the Timing Analyzer HP 16510B
7 - 2 Front-Panel Reference
Page view 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 271 272