6. Error Signals
Error Possible Cause Elimination
Transmitter does not respond No supply voltage Connect data cable or power pack
Incorrect serial interface type Remove rubber stoppers on rating
or incorrect address set plate side and set address switch to
the serial interface to be used
and/or set the correct address
(see Section 3.3.).
Transmitter defective Replace transmitter
A too high measuring value is High transmitter degassing Degas (see Section 5.)
printed out in the high vacuum
Adjustment necessary, Re-adjust (see Section 4.1.)
contamination
Transmitter defective Replace transmitter
Deposits on the electrode of the hot cathode tubes can cause instability in the
measurement signals and lead also to increased desorption in the ultra high vacuum range
which, in turn, can lead to a too high measured pressure.
In such cases it is desirable, at pressures of below 1.5 x 10
-6
Torr (2 x 10
-6
mbar), to clean the
anode of the transmitter by baking out the adsorbed gas particles. This is carried out (via the
connected measuring unit) by starting the Windows Software DokuStar in accordance with
the
communication instructions relating to the degassing procedure. In this way the anode is
warmed up by ohmic heat up to 800 °C (1500°F). Although baking out lasts approximately 3
minutes, the process can be discontinued at any time beforehand.
No transmission of measurement values is possible during the time the transmitter is in
degassification mode.